FP414 : THE SIMILARITY IN THE DISSIMILAR – Comparison of macular thickness using 3 different SD-OCT systems.

  1. Home
  2. /
  3. Poster Podium Presentation, Diabetic Retinopathy & Medical...
  4. /
  5. FP414 : THE SIMILARITY...

FP414 : THE SIMILARITY IN THE DISSIMILAR – Comparison of macular thickness using 3 different SD-OCT systems.

Share this post

Dr.Pushpanjali, R16749

Aim:To compare foveal,parafoveal and perifoveal thickness using 3 different SD-OCT systems.

Methods:It was a prospective study where normal subjects age between 20-50 years were included.Subjects

with any ocular surgery, systemic and retinal disease were excluded.All subjects underwent foveal,parafoveal thickness and perifoveal thickness measurement using Optovue OCT,Optovue OCT Angio

and Zeiss OCT.

Results:40 eyes of 20 subjects with mean±Sd age 22.2±1.66 years were included.The foveal and

perifoveal thickness was similar among all three instruments(p>0.05).Parafoveal thickness measured using 3 instruments was similar except superior region (p<0.05).Superior parafoveal thickness was statistically

significantly lower with Zeiss OCT as compared to Optovue OCT Angio however this difference was

clinically insignificant.

Conclusion:Foveal,parafoveal and perifoveal thickness measurements using 3 different OCT systems were similar and these machines can be used interchangeably by clinicians.