Dr.Unnikrishnan Nair R, Dr.Manoj S
Small gauge MIVS especially 27 G MIVS has been catching on albeit a tendency to consider it unsuitable for complicated cases and limit its use to simpler vitrectomies associated with macular pathology and vitreous removal. However one may consider the real effectiveness and beauty of the vitrectomy cutter is the ability to modulate its parameters to work as many instruments in one form. An understanding of duty cycle , cut rates , vacuum control alteration may be used effectively to deal with even difficult cases including diabetic membranes. The cutter appropriately configured may be used as a forceps, scissors, aspirator , delaminator and can function as a single multipurpose instrument without resorting to multiple change overs during surgery. This video with examples will demonstrate how to understand the machine capability and apply this knowledge to effectively improve your surgical skills while dealing with difficult vitreo-retinal cases